Local Coverage Determination (LCD)

BDX-XL2

L37216

Expand All | Collapse All
Proposed LCD
Proposed LCDs are works in progress that are available on the Medicare Coverage Database site for public review. Proposed LCDs are not necessarily a reflection of the current policies or practices of the contractor.

Document Note

Note History

Contractor Information

LCD Information

Document Information

Source LCD ID
N/A
LCD ID
L37216
Original ICD-9 LCD ID
Not Applicable
LCD Title
BDX-XL2
Proposed LCD in Comment Period
N/A
Source Proposed LCD
DL37216
Original Effective Date
For services performed on or after 09/16/2017
Revision Effective Date
For services performed on or after 03/30/2023
Revision Ending Date
N/A
Retirement Date
N/A
Notice Period Start Date
08/01/2017
Notice Period End Date
09/15/2017
AMA CPT / ADA CDT / AHA NUBC Copyright Statement

CPT codes, descriptions and other data only are copyright 2023 American Medical Association. All Rights Reserved. Applicable FARS/HHSARS apply.

Fee schedules, relative value units, conversion factors and/or related components are not assigned by the AMA, are not part of CPT, and the AMA is not recommending their use. The AMA does not directly or indirectly practice medicine or dispense medical services. The AMA assumes no liability for data contained or not contained herein.

Current Dental Terminology © 2023 American Dental Association. All rights reserved.

Copyright © 2023, the American Hospital Association, Chicago, Illinois. Reproduced with permission. No portion of the American Hospital Association (AHA) copyrighted materials contained within this publication may be copied without the express written consent of the AHA. AHA copyrighted materials including the UB‐04 codes and descriptions may not be removed, copied, or utilized within any software, product, service, solution or derivative work without the written consent of the AHA. If an entity wishes to utilize any AHA materials, please contact the AHA at 312‐893‐6816.

Making copies or utilizing the content of the UB‐04 Manual, including the codes and/or descriptions, for internal purposes, resale and/or to be used in any product or publication; creating any modified or derivative work of the UB‐04 Manual and/or codes and descriptions; and/or making any commercial use of UB‐04 Manual or any portion thereof, including the codes and/or descriptions, is only authorized with an express license from the American Hospital Association. The American Hospital Association (the "AHA") has not reviewed, and is not responsible for, the completeness or accuracy of any information contained in this material, nor was the AHA or any of its affiliates, involved in the preparation of this material, or the analysis of information provided in the material. The views and/or positions presented in the material do not necessarily represent the views of the AHA. CMS and its products and services are not endorsed by the AHA or any of its affiliates.

Issue

Issue Description

Review completed 03/02/2023 with no change in coverage.

Issue - Explanation of Change Between Proposed LCD and Final LCD

CMS National Coverage Policy

Title XVIII of the Social Security Act (the “Act”), Section 1862(a)(1)(A). This section limits coverage and payment to those items and services that are reasonable and necessary for the diagnosis or treatment of illness or injury or to improve the functioning of a malformed body member.

42 FR 410.32 (a) Diagnostic x-ray tests, diagnostic laboratory tests, and other diagnostic tests: Conditions

CMS Internet-Only Manual, Pub. 100-02, Medicare Benefit Policy Manual, Chapter 15, §80 Requirements for Diagnostic X-Ray, Diagnostic Laboratory, and Other Diagnostic Tests and §80.1.1 Certification Changes

Coverage Guidance

Coverage Indications, Limitations, and/or Medical Necessity

This Medicare contractor will provide limited coverage for the (BDX-XL2) test (Biodesix, Boulder, CO and Seattle, WA) for the management of a lung nodule, between 8 and 30mm in diameter, in patients 40 years or older, and with a pre-test cancer risk (as assessed by the Mayo Clinic Model for Solitary Pulmonary Nodules) of 50% or less. The intended use of the test is to assist physicians in the management of lung nodules by identifying those lung nodules with a high probability of being benign. These lung nodules would then be candidates for non-invasive computed tomography (CT) surveillance instead of invasive procedures.

Summary of Evidence

Lung nodules are rounded densities, under 30mm in diameter, detected by chest radiograph (CXR) or CT scan. Nodules are mostly surrounded by lung tissue and are also called coin lesions, solitary pulmonary nodules or lesions, or a “spot” on the lung. The edges of a nodule can be described as smooth or irregular (stellate or spiculated) with irregular edges somewhat more indicative for cancer. Heavily calcified nodules with smooth edges are generally benign and solid nodules that have not shown growth over time are considered benign. For a comprehensive review of nodules and their evaluation, see the 2-part series by Patel, et al.1,2

Lung nodules are detected incidentally or through lung cancer screening. Lung cancer screening now has Medicare coverage in the United States (US). As of April 26, 2016, there were 806 sites registered for screening.22 The estimated number of new lung nodules incidentally detected annually in the US is 1.57 million, whereas it is anticipated that another 1.5 million are detected by screening annually in the US.3

Early detection of lung nodules is an opportunity to reduce lung cancer mortality, but it comes with significant risks. These risks are both for patients and health care delivery networks. For patients, a major problem is the risk of unnecessary invasive procedures to find the minority of nodules that are cancer. For health care delivery, the risks are both costs and overloading the health care system.

Of the expected 3 million nodules per year found incidentally or by lung cancer screening, the majority will be Medicare age. Four recent studies underscore the importance of lung cancer evaluations to the Medicare population:

  • The mean age of 377 eligible patients in an 18-site retrospective chart review study was 65.4
  • A prospective study across 12 sites and 475 patients found 62.5% of patients were 65 years of age or older.5
  • A recent study reported that between Jan 2009 and Dec 2011, 8,979 Medicare patients from a random sampling of 5% of Medicare claims, underwent lung cancer evaluations because of an abnormal chest CT scan.6
  • In the National Lung Screening Trial (NLST), over 57% of enrollees were over 65 years of age.7 Also, the rates of nodule detection increased dramatically with age.3 Medicare enrollees are more likely to meet lung cancer screening criteria and have more nodules detected.

A set of guidelines for lung nodule management is published and updated by the American College of Chest Physicians (ACCP). The ACCP guidelines for lung nodules, updated in 2013, is the primary reference used by pulmonologists in the US.8 The ACCP Guidelines state: “Although clinical and radiographic [CT scans] characteristics cannot reliably distinguish between benign and malignant nodules in most individuals, it is nevertheless important to estimate the clinical probability of malignancy before ordering imaging tests or biopsy procedures”. The pretest probability of malignancy (pCA) is estimated by using clinical judgment or with a quantitative risk model.9,10 Establishing a pCA creates 3 risk stratification groups, namely, Low, Intermediate, and High probability, with Low risk having pCA below 5% and High risk having pCA above 65%. The general concept is that Low risk patients will be observed with CT surveillance to watch for growth if a nodule is malignant. Conversely, the guidelines suggest those patients in the High-risk group go directly to surgery. The logic is that the probability of cancer is high enough that a negative biopsy will not change the care pathway. The Intermediate risk group (5-65% pCA) are recommended to enter the diagnostic odyssey that often includes positron emission tomography (PET) scanning as the next step. A negative PET suggests a benign nodule, so the patient is followed with CT scans. A positive PET scan goes on to surgery or biopsy. This is the overall concept, but PET has sensitivity and specificity challenges. In particular, current estimates of PET sensitivity from 72% to 94% are reviewed in the 2013 ACCP Guidelines in section 4.2.3.8 False positive PET scans for nodules are an additional problem with estimates of the false positive PET scan rate of 39%.4

A pulmonary community practice observational chart review of 18 practices and 377 patients found a wide variation in management of nodules.4 The surgery rate for benign nodules was 35%, and the rate of surgery was the same for Low, Intermediate, and High risk patients. The risk categories were calculated by the study and despite a Low risk, 28% had biopsies and 17% had surgery. The rates of surgery for benign nodules range between 10% and 55%. A survey of 196 pulmonologists supports the potential of a non-invasive biomarker to positively improve lung nodule management decisions.11

Biopsies can be obtained through a bronchoscope, or a needle passed through the chest wall with CT image guidance. A community practice chart review found 38% of patients had a form of biopsy.4 Complications with biopsies or surgery are increased with age, smoking history, and other lung disease. Biopsy through the bronchoscope has the lowest risk with a 2-4% risk of bleeding or pneumothorax.8 A disadvantage of this procedure is inaccurate sampling of the nodule. Correct sampling averages about 50%.8 The correct sampling rate may improve with modern navigation techniques that are being adopted. Bronchoscopic biopsy use for nodules is currently about 20% of nodules.6 Needle biopsies are done in about 15% of patients with nodules with a 1% risk of bleeding, and a 15-19% risk of pneumothorax.12 About half (7%) of patients with a pneumothorax require chest tube placement with a significant period of hospitalization.12 Most needle biopsies are diagnostic, but the risk of a non-diagnostic result with a malignant nodule is about 20%.8 Biopsies (combined bronchoscopy and needle) are performed in about 25% of nodules (200,000), and the procedures are for benign nodules in 42-62% (104,000). Complications from biopsies result in hospitalization in 2-7% of cases.6,12 That translates into 4,680 excess hospitalizations per year that are potentially avoidable.

Eventually, most malignant nodules go to surgery for resection and about 15-25% of patients have biopsy attempts before surgery. The overall surgery rate is about 34% (270,000 per year) for benign and malignant nodules in the nodule population.4 Complications include death (2% in Medicare population),13 prolonged lung air leak (3-5%), and pneumonia (1-8%).8 Published rates for surgery for benign nodules range from 31-44%.4,7,17 This translates into an estimated 102,000 surgeries and 2,052 deaths per year that are avoidable for patients that do not have lung cancer.

Test Description and Intended Use
BDX-XL2 is a proteomic risk predictor that integrates the expression levels of 2 proteins with 5 clinical risk factors. The BDX-XL2 assay is performed on fresh-frozen EDTA plasma samples using mass spectrometry.15,16 Results are reported as “Likely Benign” when the post-test probability that a lung nodule is benign is 90% or higher. Otherwise, the test reports “Indeterminate” when the post-test probability is less than 90%. “Likely Benign” test reports also include the post-test probability that the lung nodule is benign, ranging from 90% to 98%. This is further detailed in Table 1 below, along with the performance of the test at each post-test probability.

The intended use of the test is to assist physicians in the management of lung nodules by identifying those lung nodules with a high post-test probability of being benign. These lung nodules would then be candidates for non-invasive CT surveillance instead of invasive diagnostic procedures such as biopsy or surgery.

Clinical Validation
The discovery, clinical validation, and analytical performance of earlier versions of the assay were previously reported.15-17 The current version of (BDX-XL2) is a refinement that incorporates clinical risk factors (nodule size, age, smoking history, nodule location, and nodule spiculation). BDX-XL2 was retrospectively validated on the prospective observational PANOPTIC study (NCT01752114) of lung nodule management. PANOPTIC enrolled 685 subjects across 33 sites in the US and Canada. Validation of BDX-XL2 followed the National Academy of Medicine’s guidelines for rigorous test development.18

In the PANOPTIC study, 178 subjects met the intended use population of BDX-XL2. This consisted of 149 benign lung nodules (as determined by histopathology after biopsy or surgery, or by stable CT surveillance of a lung nodule after at least 1 year) and 29 malignant lung nodules (as determined by histopathology after biopsy or surgery), yielding a cancer prevalence of 16.3%. Per protocol, this cohort of patients was split into separate verification (n = 69) and validation (n = 109) subsets, with pre-specified interim and final analyses on each subset, respectively, to determine the performance characteristics of the test. Since the test system and clinical endpoints were unchanged between the interim and final analyses, all analyses were blinded, and the 2 patient cohorts were mutually exclusive; the 69 and 109 patients are combined below.

For these 178 subjects, BDX- XL2 yielded a sensitivity, specificity, negative predictive value (NPV) and positive predictive value (PPV) of 97% (CI: 82%-100%), 44% (CI: 36%-52%), 98% (CI: 92%-100%), and 25% (CI: 17%-34%), respectively. We note that the post-test probability of BDX-XL2 is equivalent to its NPV.

Procedure use and clinical factors were collected in the PANOPTIC study permitting the comparison of BDX-XL2 performance to:

  • current practice, as characterized by physician pre-test risk assessment, denoted as pCA, which is based on the physician’s clinical assessment of risk using a clinical risk model, and/or clinical judgement,
  • PET, and
  • clinical risk factor models including the ‘Mayo’, ‘VA’, ‘Brock’, and ‘Herder’ models (9, 10, 19, 20).

Using area under the curve (AUC) as a general performance measure, the respective AUCs for XL2, pCA, PET, and these 4 clinical factor models were: 76% (CI: 69%-82%), 69% (CI: 62% - 76%), 58% (CI: 46%-69%), 69% (CI: 62%-76%), 60% (CI: 53%-67%), 71% (CI: 63%-77%), and 67% (CI: 56%-78%). BDX-XL2 had statistically significant improved performance over all above diagnostic modalities with p-values 2.1xE-11 (pCA), 0.001 (PET), 0.0009 (Mayo), 2.7xE-7 (VA), 0.005 (Brock) and 0.02 (Herder).

Clinical Utility
Current practice has been previously characterized,4,11 as well as, the potential clinical utility of an earlier version.5 The clinical utility of BDX-XL2 is measured in terms of its potential to reduce unnecessary invasive procedures, such as biopsies and surgeries, on benign lung nodules while not significantly increasing the number of malignant lung nodules routed to CT surveillance, thereby delaying surgical resection. Assuming strict adherence to management recommendations based on assay results (i.e., active surveillance if “likely benign”), an earlier version of Xpresys® Lung demonstrated a potential 32% reduction in invasive procedures on benign lung nodules without increasing the number of malignant nodules routed to CT surveillance5 based on a retrospective analysis of a prospective observational study of lung nodule management (NCT01752101). Similarly, in the PANOPTIC study, the potential clinical utility of the current version of (BDX-XL2) was assessed retrospectively. Specifically, if BDX-XL2 were used to guide lung management (and assuming a post-test probability of 98%), then invasive procedures on benign lung nodules would have been reduced 36% (CI: 22%-52%) with only 3% (CI: 0%-18%) of malignant nodules routed to CT surveillance (compared to 45% with current practice in the PANOPTIC study).

Summary of Analytical and Clinical Performance
General

Intended Use BDX-XL2 is intended for the evaluation of 8-30 mm lung nodules in patients 40 years or older with a pre-test cancer risk (as assessed by the Mayo Clinic Model for Solitary Pulmonary Nodules) of 50% or less.
Validated Specimen Type(s) Plasma from K2 EDTA vacutainer tubes

 

Analytical Performance

Description Results
(with 95% Confidence Intervals if applicable)

Repeatability (within run precision)

8 samples tested 3 times within a run, 1 instrument, 2 operators, 3 runs, 3 non-consecutive days, 1 manufacturing reagent lot for critical reagents. Qualitative results represent only 2 possible final results (i.e., indeterminate and likely benign) with post-test probabilities for the latter from 90 to 98%

Analytes (analytical repeatability):
ARRLG3BP CV = 7.0% (3.6%-10.3%)
ARRC163A CV = 5.9% (3.6%-8.2%)

Score (analytical repeatability):
Score CV = 9.0% (3.4%-14.6%)

Qualitative (clinical concordance):
62.5% (5/8; 24.5%-91.5%)

ARR is defined as the ratio between the peak area for the endogenous quantifier peak and the peak area of the corresponding SIS peptide, multiplied by a calibration factor that is specific for each SIS lot.

Score is defined as Log2[ARRLG3BP/ ARRC163A]
Intermediate precision (between run precision)

7 samples tested once in 3 different runs, 1 instrument, 2 operators, 3 runs, 3 non-consecutive days, 1 manufacturing reagent lot for critical reagents. Qualitative results represent only 2 possible final results (i.e., indeterminate and likely benign) with post-test probabilities for the latter from 90 to 98%.
Analytes (analytical repeatability):
ARRLG3BP CV = 14.5% (10.1%-18.9%)
ARRC163A CV = 12.8% (9.1%-16.6%)

Score (analytical repeatability):
Score CV = 6.0% (3.7%-8.4%)

Qualitative (clinical concordance):
85.7% (6/7; 42.1%-99.6%)
Reproducibility (between sites) Not applicable
Minimum input quantity 20 μL plasma
Limit of blank (LOB) Defined as the upper 99.5% confidence interval (CI) for the response ratio (ARR) observed in negative controls.
Limit of detection (LOD) The lower limit of response for each analyte in samples and positive controls is defined as the lower limit of quantification.
Limits of quantitation (LOQ) The lower limit of quantification (LLOQ) is defined as the lowest response ratio within the linear range where the coefficient of variation (CV) was equal to or below 0.20, where linearity was established using 5 replicate measures at each concentration.

The upper limit of quantification (ULOQ) is defined as the highest response ratio within the linear range where the coefficient of variation (CV) was equal to or below 0.20, where linearity was established using 5 replicate measures at each concentration.

LG3BP:
ULOQ = 42
LLOQ = 0.043

C163A:
ULOQ = 71
LLOQ = 0.053
Linearity Not applicable for qualitative interpretation (i.e., likely benign or indeterminate)
For individual analytes, the linear response range was established between the LLOQ and ULOQ
Interfering substances Visual inspection to detect hemolysis (100 mg/dL of hemoglobin rejected).

MRM-MS chromatograms for each analyte in every sample are inspected for interference, with any interference resulting in QC failure.
Specimen stability, primary (EDTA whole blood) 3 hours at 2-8 °C based on validation study (manuscript in preparation)
Specimen stability, intermediate (plasma) 24 months at -70 °C based on validation study (manuscript in preparation)

2 freeze-thaw cycles when stored at -70 °C based on validation study (manuscript in preparation)
Reagent closed/shelf-life stability 24 months at -70 °C for 2 critical reagents (i.e., Human Plasma Samples (HPS) and SIS peptides). ARR for 623 samples evaluated with slope not significantly different from zero (p = 0.77).

Non-critical (general purpose) reagents are stored and expired per manufacturer recommendations.
Reagent open/in use stability Critical reagents: Not applicable since single use aliquots

Non-critical reagents are stored and expired per manufacturer recommendations.

 

Clinical Performance: Validity
BDX-XL2 integrates the relative abundance of 2 plasma proteins (LG3BP and C163A) with 5 clinical risk factors (age, smoking status, nodule diameter, nodule spiculation status and nodule location). From these 7 markers, a numerical value, XL_2(k), for a patient k, is calculated.

XL_2(k) ranges between 0 and 1 and its value is used to index the post-test probability (i.e., NPV) validated in the PANOPTIC study (see Table 1 below).

Table 1: XL2 Performance and Test Reporting (n= 178 subjects).

XL_2(k)Value Post-Test
Benign
Probability
(i.e. NPV)
(95% CI)
Sensitivity
(95% CI)
Specificity
(95% CI)
PPV
(95% CI)
Test
Report
0 to 0.131 98%
(92% - 100%)
97%
(82% - 100%)
44%
(36% - 52%)
25%
(17% - 34%)
Likely Benign
>0.131 to 0.1613 97%
(91% - 100%)
93%
(77% - 99%)
49%
(41% - 57%)
26%
(18% - 36%)
Likely Benign
0.1613 to 0.172 96%
(90% - 99%)
90%
(73% - 98%)
54%
(45% - 62%)
27%
(19% - 37%)
Likely Benign
>0.172 to 0.176 95%
(89% - 99%)
86%
(68% - 96%)
55%
(47% - 63%)
27%
(18% - 37%)
Likely Benign
>0.176 to 0.1785 94%
(87% - 98%)
83%
(64% - 94%)
56%
(47% - 64%)
27%
(18% - 37%)
Likely Benign
>0.1785 to 0.193 93%
(86% - 98%)
79%
(60% - 92%)
57%
(44% - 65%)
26%
(18% - 37%)
Likely Benign
>0.193 to 0.195 92%
(85% - 96%)
76%
(56% - 90%)
58%
(49% - 66%)
26%
(17% - 37%)
Likely Benign
>0.195 to 0.2306 91%
(84% - 96%)
69%
(49% - 85%)
64%
(56% - 72%)
27%
(18% - 39%)
Likely Benign
>0.2306 to 0.354 90%
(84% - 95%)
55%
(36% - 74%)
83%
(75% - 88%)
38%
(24% - 54%)
Likely Benign
>0.354 - - - - Indeterminat

 

Clinical Performance: Utility
PANOPTIC was a non-interventional study; however, the potential clinical utility of BDX-XL2 can be estimated by evaluating how many benign (benefit) and malignant (harm) nodules would have been routed away from invasive procedures into CT surveillance if BDX-XL2 had been used to guide patient management in the study (and assuming complete compliance). Table 2 summarizes the potential clinical utility of BDX-XL2 at each post-test probability.

Table 2: Potential Clinical Utility of BDX-XL2v

Post-Test
Probability
of Being Benign
Reduction in Invasive Procedures
of Benign Nodules
(95% CI)
Malignant Nodules Sent to
CT Surveillance
(95% CI)
98% 15/42 = 36%
(22% - 52%)
1/29 = 3%
(0% -18%)
97% 17/42 = 40%
(26% - 57%)
2/29 = 7%
(1% -23%)
96% 19/42 = 45%
(30% - 61%)
2/29 = 7%
(1% -23%)
95% 20/42 = 48%
(32% - 64%)
2/29 = 7%
(1% -23%)
94% 20/42 = 48%
(32% - 64%)
2/29 = 7%
(1% -23%)
93% 22/42 = 52%
(36% - 68%)
3/29 = 10%
(2% -27%)
92% 22/42 = 52%
(36% - 68%)
4/29 = 14%
(4% -32%)
91% 25/42 = 60%
(43% - 74%)
4/29 = 14%
(4% -32%)
90% 32/42 = 76%
(61% - 88%)
7/29 = 24%
(10% -44%)

 

In the PANOPTIC study, there were 178 intended use subjects (when the verification and validation sets are combined) of whom 29 had malignant lung nodules and 149 had benign lung nodules. Of the 149 benign lung nodules, 42 had at least 1 invasive procedure. Hence the denominator of “42” in column 2 of Table 2. Of the 29 malignant lung nodules, 13 were routed to CT surveillance. That is, 13/29 = 45% of malignant nodules were routed to CT surveillance in PANOPTIC. This is substantially larger than the largest corresponding value in Table 2 (i.e., 24%).

Coverage Summary
The BDX-XL2 assay is reasonable and necessary to assist physicians in the management of lung nodules by identifying those lung nodules with a high probability of being benign. This assay is only covered when the following conditions are met:

  • Patient is at least 40 years of age and has a lung nodule of diameter 8 to 30mm, and
  • The pre-test risk of cancer as determined by the Mayo risk prediction algorithm (10) is 50% or less

Note: The BDX-XL2 test should not be ordered if a physician does not intend to act upon the test result. It is expected that physicians will advise nodule surveillance for at least 80% of patients with a post-test probability of 98% or higher.

Analysis of Evidence (Rationale for Determination)

Level of evidence

Quality – Moderate
Strength – Moderate
Weight - Limited

This contractor recognizes that evidence of clinical utility for the (BDX-XL2) assay for ≥ 40-year-old patients with an 8 to 30 mm lung nodule and a pre-test cancer risk (as assessed by the Mayo Clinic Model for Solitary Pulmonary Nodules) of ≤ 50% is promising at the current time.  Clinical Studies underway at this time are expected to demonstrate clinical utility.  These studies are designed to show a statistically significant reduction in the number of benign lung nodules experiencing invasive procedures between a prospective group of patients managed by BDX-XL2 and a contemporaneous group not managed by BDX-XL2.  A secondary end-point will show that the management of lung nodules by BDX-XL2 does not (i.e., is statistically non -inferior to) the number of malignant nodules routed to CT surveillance (determined at 1-year interval) as compared to current practice without BDX-XL2.  Continued coverage for BDX-XL2 testing will be dependent on annual review of prospective data and peer-reviewed studies.

Data collected by Biodesix through ongoing studies will support utility including:

• All clinical risk factors to calculate the Mayo, VA, and Brock cancer risk predictors
• PET result (if used)
• Physician-assessed pre-test cancer risk assessment
• Physician post-test lung nodule management recommendation
• Any subsequent procedures (non-invasive or invasive)
• Clinical diagnosis based on those procedures (i.e., benign or malignant)

Proposed Process Information

Synopsis of Changes
Changes Fields Changed
N/A
Associated Information
Sources of Information
Bibliography
Open Meetings
Meeting Date Meeting States Meeting Information
N/A
Contractor Advisory Committee (CAC) Meetings
Meeting Date Meeting States Meeting Information
N/A
MAC Meeting Information URLs
N/A
Proposed LCD Posting Date
Comment Period Start Date
Comment Period End Date
Reason for Proposed LCD
Requestor Information
This request was MAC initiated.
Requestor Name Requestor Letter
N/A
Contact for Comments on Proposed LCD

Coding Information

Bill Type Codes

Code Description
N/A

Revenue Codes

Code Description
N/A

CPT/HCPCS Codes

Group 1

Group 1 Paragraph

N/A

Group 1 Codes

N/A

N/A

ICD-10-CM Codes that Support Medical Necessity

Group 1

Group 1 Paragraph:

N/A

Group 1 Codes:

N/A

N/A

ICD-10-CM Codes that DO NOT Support Medical Necessity

Group 1

Group 1 Paragraph:

N/A

Group 1 Codes:

N/A

N/A

Additional ICD-10 Information

General Information

Associated Information

N/A

Sources of Information

N/A

Bibliography
  1. Patel VK, Sagar KN, Naidich DP, et al. A practical algorithmic approach to the diagnosis and management of solitary pulmonary nodules. Part 2: Pretest probability and algorithm. Chest. 2013;143(3):840-846.
  2. Patel VK, Sagar KN, Naidich DP, et al. A practical algorithmic approach to the diagnosis and management of solitary pulmonary nodules. Part 1: Radiologic characteristics and imaging modalities. Chest. 2013;143(3):825-839.
  3. Gould MK, Tang T, Liu IA, et al. Recent Trends in the identification of incidental pulmonary nodules. American Journal of Respiratory and Critical Care Medicine. 2015;192(10):1208-1214.
  4. Tanner NT, Aggarwal J, Gould MK, et al. Management of pulmonary nodules by community pulmonologists: A multicenter observational study. Chest. 2015;148(6):1405-1414.
  5. Vachani A, Hammoud Z, Springmeyer S, et al. Clinical utility of a plasma protein classifier for indeterminate lung nodules. Lung. 2015;193:1023-1027.
  6. Lokhandwala T, Bittoni MA, Dann RA, et al. Costs of diagnostic assessment for lung cancer: A medicare claims analysis. Clin Lung Cancer. 2016;18(1):e27-e34.
  7. The National Lung Screening Trial Research. Reduced lung-cancer mortality with low-dose computed tomographic screening. N Engl J Med. 2011;365(5):395.
  8. Gould MK, Donington J, Lynch WR, et al. Evaluation of individuals with pulmonary nodules: When is it lung cancer? Chest. 2013;143(5):e93S-e120S.
  9. Gould MK, Ananth L, Barnett PG. A clinical model to estimate the pretest probability of lung cancer in patients with solitary pulmonary nodules. Chest 2007;131(2):383-388.
  10. Swensen SJ, Silverstein MD, Ilstup DM, Schleck CD, Edell ES. The probability of malignancy in solitary pulmonary nodules: Application to small radiologically indeterminate nodules. Arch Intern Med. 1997;157:849-855.
  11. Vachani A, Tanner NT, Aggarwal J, et al. Factors that influence physician decision making for indeterminate pulmonary nodules. Ann Am Thorac Soc. 2014;11(10):1586-1591.
  12. Weiner RS, Schwartz LM, Woloshin S, Welch HG. Population-based risk for complications following transthoracic needle lung biopsy of a pulmonary nodule. Ann Intern Med. 2011;155(3):137-144.
  13. Iniguez CE, Armstrong KW, Cooper Z, et al. Thirty-Day mortality after lobectomy in elderly patients eligible for lung cancer screening. Ann Thorac Surg.2016;101:541-6.
  14. Wiener RS, Gould MK, Slatore CG, Fincke BG, Schwartz LM, Woloshin S. Resource use and guideline concordance in evaluation of pulmonary nodules for cancer: Too much and too little care. JAMA Intern Med. 2014;174(6):871-880.
  15. Li X, Lee LW, Brusniak M, et al. An integrated quantification method to increase the precision, robustness, and resolution of protein measurement in human plasma samples. Clinical Proteomics. 2015;12(3):1-17.
  16. Vachani A, Pass HI, Rom WN, et al. Validation of a multiprotein plasma classifier to identify benign lung nodules. Journal of Thoracic Oncology. 2015;10(4):629-637.
  17. Li X, Hayward C, Fong P, et al. A blood-based proteomic classifier for the molecular characterization of pulmonary nodules. Sci Transl Med. 2013;5(207):1-24.
  18. Silvestri GA, Tanner NT, Kearnery P, et al. Assessment of plasma proteomics biomarker’s ability to distinguish benign from malignant lung nodules: Results of the PANOPTIC (Pulmonary Nodule Plasma Proteomic Classifier) Trial. Chest. 2018;154(3):491-500.
  19. Micheel CM, Nass SJ, Omenn GS. Evolution of Translational Omics: Lessons Learned and the Path Forward. Washington, DC: National Academies Press; 2012.
  20. McWilliams A, Tammemagi MC, Mayo JR, et al. Probability of cancer in pulmonary nodules detected on first screening CT. N Engl J Med. 2013;369(10):910-919.
  21. Herder GJ, Tinteren H, Golding RP, et al. Clinical prediction model to characterize pulmonary nodules: Validation and added value of 18F-fluorodeoxyglucose positron emission tomography. Chest. 2005;128:2490-2496.
  22. American College of Radiology Online. Lung Cancer Screening Registry-Core Documents Web site. Available at: https://www.acr.org/Practice-Management-Quality-Informatics/Registries/Lung-Cancer-Screening-Registry. Accessed February 27, 2023.

Revision History Information

Revision History Date Revision History Number Revision History Explanation Reasons for Change
03/30/2023 R10

Posted 03/30/2023: Review completed 03/02/2023 with no change in coverage. Punctuation and grammatical corrections made throughout the LCD. Accession date updated in reference number 22 of the bibliography.

  • Other (Review)
09/30/2021 R9

09/30/2021-Removed “MolDX:” from the title of this LCD.

  • Other
04/22/2021 R8

04/22/2021 Under CMS National Coverage Policy added regulation CMS Internet-Only Manual, Pub. 100-02, Medicare Benefit Policy Manual, Chapter 15, §80 Requirements for Diagnostic X-Ray, Diagnostic Laboratory, and Other Diagnostic Tests and §80.1.1 Certification Changes. Under Bibliography changes were made to citations to reflect AMA citation guidelines. Registered marks were added throughout the LCD where applicable. Formatting, punctuation and typographical errors were corrected throughout the LCD. Acronyms were defined and inserted where appropriate throughout the LCD. Reviewed 03/02/2021.

  • Provider Education/Guidance
11/19/2020 R7

11/19/2020 Consistent with the MolDX contractor, under CMS National Coverage Policy: moved references to CMS Internet-Only Manual, Pub 100-02, Medicare Benefit Policy Manual, Ch. 15, added §80.0, §80.1 and §80.2. and CMS Internet-Only Manual, Pub 100-04, Medicare Claims Processing Manual, Ch. 16. §60.1 and §60.2 to A57558 Billing and Coding: MolDX; BDX-XL2. Under Summary of Evidence, removed the hyperlink http://www.acr.org/quality-safety/national-radiology-data-registry/lung-cancer-screening-registry – see Core Documents, accessed June 20, 2016 and added it to the Bibliography as citation #22. Updated bibliography to AMA guidelines. Formatting, punctuation and typographical errors corrected throughout the LCD. Acronyms defined where appropriate. Review completed 10/12/2020.

  • Other (review)
11/28/2019 R6

03/26/2020 Review completed 02/14/2020. Corrected typo.

  • Other (review)
11/28/2019 R5

11/28/2019 Change Request (CR) 10901 Local Coverage Determinations (LCDs): it will no longer be appropriate to include Current Procedure Terminology (CPT)/Health Care Procedure Coding System (HCPCS) codes or International Classification of Diseases Tenth Revision-Clinical Modification (ICD-10-CM) codes in the LCDs. All CPT/HCPCS and ICD-10 codes have been removed from this LCD and placed in Billing and Coding: MolDX: BDX-XL2 related to this LCD. Under ICD-10 Codes that Support Medical Necessity added language to Paragraph 2 and ICD-10 code R91.8 to Table 2; added modifier KX to Modifier Table in the related Article. Revised references under CMS National Coverage Policy section.

  • Other (compliance with CR 10901)
01/01/2019 R4

03/28/2019-Removed 81599 in CPT/HCPCS Group 1 and added 0080U. This revision is due to the 2019 Q1 CPT/HCPCS Updates and is effective 01/01/2019.

  • Revisions Due To CPT/HCPCS Code Changes
01/01/2019 R3

01/01/2019-Changed the title of the LCD to BDX-XL2 and all Xpresys references to BDX-XL2. Removed “The test is ordered by a physician certified in the XL2 Certification and Training Registry (CTR), and The following information is recorded: all clinical risk factors to calculate the Mayo, VA, and Brock cancer risk predictors; PET result (if used), physician pre-test risk assessment, physician post-test lung nodule management recommendation, any subsequent procedures (non-invasive or invasive), and clinical diagnosis based on those procedures (i.e., benign or malignant)” from the coverage summary section. Added additional data collected by Biodesix information to the Analysis of Evidence section. Changed the Level of evidence strength from Limited to Moderate. Added #18 to bibliography.

  • Other
07/01/2018 R2

 

07/01/2018-Annual Review completed 06/06/2018. Typo corrected.

  • Other (Annual Review)
09/16/2017 R1

07/28/2017 - In Summary of Evidence: Paragraph 5 superscripted the number 8. In the second line in the paragraph directly above Test Description and Intended Use: superscripted the number 4. In the first line under Clinical Validation, superscripted numbers 15-17. In the fifth sentence under Clinical Utility, superscripted the number 5. In Analytical Performance table, Row 9 Column 2: removed &;ge.

  • Typographical Error
N/A

Associated Documents

Attachments
N/A
Related National Coverage Documents
N/A
Public Versions
Updated On Effective Dates Status
03/22/2023 03/30/2023 - N/A Currently in Effect You are here
Some older versions have been archived. Please visit the MCD Archive Site to retrieve them.

Keywords

N/A

Read the LCD Disclaimer